Specifications E1
| Sensors and sample handling system |
|
| Number of sensors |
1 |
| Volume above sensor |
40 µl |
| Typical sample volume |
~ 500 µl |
| Working temperature |
15-65 ° C, controlled via software |
| Temperature stability |
± 0.02 ° C |
| Flow rates |
0-1 ml/min |
| Cleaning |
All parts easily removed for separate cleaning |
|
|
| Frequency and dissipation characteristics |
|
| Frequency range |
1-70 MHz |
| Maximum measurement rate |
Up to 200 data points per second |
| Typical mass sensitivity in liquid |
1.8 ng / cm2 (18pg/mm2) |
| Typical detectable change in D in liquid * |
0.1 x 10-6 |
|
|
| Software capabilities |
|
| General |
Plots, data editing and calculations |
Input data, analysis software
|
Multiple frequency and dissipation data |
|
Import/Export
|
Excel, BMP, JPG, WMF, etc. |
|
Ouput data, analysis software
|
Sauerbrey: Mass and thickness vs. time
Voigt: Modeled values of viscosity, elasticity, thickness vs. time
Other functions: Kinetic constants and own models
|
|
Other
|
Temperature ramps
|
| PC requirements |
USB 2.0, Windows 2000, XP |
|
|
| Dimensions |
|
| Electronics unit |
18 x 36 x 21 cm (HxWxD), 9 kg |
| Measurement chamber |
5 x 10 x 15 cm (HxWxD), 1 kg |
Specifications subject to change without notice.

For more information on our applications, products and the QCM-D technology, you are always welcome to contact us.