SPECIFICATIONS Q-SENSE E4 AUTO

Sensors and sample handling system  
Number of sensors 4, also possible using only 1, 2 or 3 sensors 
Flow profile Parallel plate flow
Typical sample volume 500 µl per sensor
Volume above sensor 40 µl
Working temperature 18-50 °C
Temperature stability  ± 0.02 °C
 Flow rates  0-800 µl/min
 Cleaning All parts easily removed for separate cleaning
   
Frequency and dissipation characteristics  
Frequency range 1-70 MHz
Maximum measurement rate: Up to 200 data points / s
Typical mass sensitivity in liquid*: 1.8 ng / cm2 (0.1 Hz)
Typical detectable change in D in liquid * 0.1 x 10-6
   
Software capabilities  
General

Plots, data editing and calculations.
Allows intuitive programming of measurements sequences using the E4 and QSoft

Ouput data, Sauerbrey Mass & thickness vs time
Output data, Voigt viscoelastic model Thickness, viscosity & elasticity vs time
Other functions Kinetic constants, temperature ramps, own models
PC requirements USB 2.0, XP, Vista
   
Autosampler Liquid         handling capabilities  
Sampler type X/Y/Z with stationary racks, arm with 4 probes 
Operation Probes connected to QCM-D flow module inlets, outlets connected to pump

Arm speed

350 mm/s in X and Y dimensions, 125 mm/s in Z dimensions
Number of samples Standard set-up 80 (20 rows x 4) for 9 ml tubes + 44 (11 rows x 4) for 20 ml tubes
   
Dimensions  
Electronics unit 18 x 36 x 21 cm (HxWxD), 9 kg
Measurement chamber 12 x 23 x 34 cm (HxWxD), 8 kg
Autosampler 57 x 60 x 54 cm (HxWxD), 22 kg